PROTOTYPES OF NIF NEUTRON TIME-OF-FLIGHT DETECTORS TESTED ON OMEGA

Prototypes of NIF Neutron Time-of-Flight Detectors Tested on OMEGA Introduction The National Ignition Facility1 (NIF) is a 2-MJ, 192-beam laser system currently under construction at Lawrence Livermore National Laboratory. One of the main missions of the NIF is to achieve thermonuclear ignition of fusion fuel in inertial confinement fusion (ICF).2 In ICF experiments, primary neutrons are produced in two reactions:

D + D → n(2.45 MeV)+ 3 He, D + T → n(14.1 MeV) + α .

(1)

(2)

Neutrons from reaction (1) are referred to as DD neutrons and neutrons from reaction (2) are referred to as DT neutrons. Every large ICF laser facility, including Nova, OMEGA, and GEKKO, uses neutron time-of-flight (nTOF) systems to measure neutron yields and ion temperatures. Such nTOF systems are usually based on current-mode detectors consisting of a fast plastic scintillator optically coupled to a fast photomultiplier tube (PMT). A high-bandwidth transient digitizer records the signal. These systems, which are relatively inexpensive, have a large dynamic range and a fast time response. The nTOF detectors are reliable, and the information they record is fundamental to most ICF implosion experiments. It is for these reasons that the nTOF system was identified as a “core” diagnostic3 for the NIF. The 30-kJ, 60-beam OMEGA laser system4 is currently the only facility that produces sufficient ICF neutrons for developing and testing prototype nTOF detectors for the NIF. Several nTOF detector prototypes have been built and tested on OMEGA. Based on the results of these tests, a set of nTOF detectors is proposed for use on the NIF to measure ion temperature and DD and DT neutron yields from 109 to 1019.

202

NIF nTOF Detector Requirements The nTOF system for the NIF must measure ion temperatures of implosion targets, ranging from 1 to 50 keV for yields between 109 and 1019 neutrons. The nTOF system must work in a harsh environment5 of energetic neutrons, x rays, γ rays, and high electromagnetic pulse (EMP) noise. The main objective of an nTOF system is to measure ion temperature. Because nTOF-detector signals are proportional to the number of neutrons detected, they are easily adapted for simultaneous use as a yield monitor. The neutrons produced in fusion reactions (1) and (2) are monoenergetic. Center-of-mass motion of the reacting ions causes spectral broadening of these energy lines. Because ICF targets are nearly ideal point sources in both time (<100 ps) and space (<100 µm) and neutrons travel to a detector without collisions, neutron spectra can be measured using the time-offlight technique. The arrival time at a detector corresponds to the energy of the neutron. The time spread ∆t of neutrons arriving at an nTOF detector is given by the following equations:6 ∆t = 0.778d Ti

(for DD),

(3)

∆t = 0.122 d Ti (for DT ),

(4)

where ∆t is the full width at half maximum (FWHM) in nanoseconds, d is target-to-detector distance in meters, and Ti is the ion temperature in keV. The width of a measured signal is the width of the time-of-flight spread added in quadrature with the detector response.7 To minimize measurement error, the nTOF detector response (FWHM) should be much less than the neutron temporal broadening being measured. For scintillator- and PMT-based detectors, this requirement leads to the use of fast microchannel-plate (MCP) PMT’s with time resolutions of a few hundred picoseconds (FWHM).

LLE Review, Volume 99

PROTOTYPES OF NIF NEUTRON TIME-OF-FLIGHT DETECTORS TESTED ON OMEGA The scintillator used in an nTOF detector must also be fast. Bicron’s8 ultrafast BC-422 scintillator has a rise time of less than 20 ps9 and an exponential decay constant of 1.4 ns. The quenched version of the scintillator, BC-422Q, has a twocomponent decay with time constants of about 0.6 and 5 ns. In addition, the nTOF detector-response function includes the transit time of the neutrons across the thickness of the scintillator. A scintillator thickness of 20 mm corresponds to a 0.92-ns transit time for DD neutrons and provides a good match between scintillator decay and transit time through the scintillator. The cables and digitizer also contribute to detector response. A cable with a 3-GHz bandwidth and a 1-GHz or faster oscilloscope contributes relatively small dispersion to temperature measurements. X rays and γ rays from an ICF implosion generate background10 in nTOF detectors that can saturate a PMT and distort the neutron signal, causing errors in ion temperature and yield measurement. Shielding is therefore required to reduce the x-ray and γ-ray fluence entering PMT-based nTOF detectors. Lead-shielding thickness is limited to <30 mm to avoid neutron scattering that would appear as signal broadening in the detector. High-yield DT implosions create MeV γ rays from (n, γ ) interactions in the target, target positioner, nearby diagnostics, and the target chamber walls. This γ-ray background is proportional to the DT neutron yield and will be very high at expected NIF yields. It is difficult to shield MeV γ rays without also shielding the energetic neutrons. New techniques10 like single-stage MCP PMT’s and chemical-vapor-deposition (CVD) diamonds,11 which are less sensitive to MeV γ rays, are recommended for the NIF. nTOF Detector Locations on the NIF The optimum placement12 of nTOF detectors is determined by a tradeoff between decreased time resolution at small distances from the target and a smaller statistical sample of detected neutrons at longer distances. It can be shown13 that the number of neutrons needed to achieve a given statistical uncertainty should be a factor of 2 larger over that predicted by Poisson statistics. Therefore, in our design, we require at least 200 neutron interactions in the nTOF detectors to achieve 10% statistical uncertainty, which puts a restriction on detector location. Another requirement is that, at the lowest yields, the nTOF detector should provide a neutron signal with an amplitude five to ten times higher than the EMP noise of the system. The highest-measurable yield is determined by PMT saturation. The signal from a modern MCP PMT is linear to ~3 nC of integrated charge for low-repetition pulses. This was taken into account in estimating yield limits. A combination of the optiLLE Review, Volume 99

mization12 of the detector locations combined with background considerations provides several natural locations for nTOF detectors on the NIF. Placing a detector outside the target chamber wall at 5 m from the target avoids vacuum interface and tritium contamination problems. A 5-m flight path is adequate for iontemperature measurements. At this location the nTOF detector will not be affected by scattered neutrons and (n,γ) interactions with the target chamber. The EMP noise at the OMEGA target chamber is 2 to 20 mV, depending on shot and detector design. The EMP noise at the NIF target chamber will most likely be higher; therefore, at least a 500-mV signal is required at the target chamber wall. This location should be used only for D2 implosions and low-yield DT shots as described in the next section and Table 99.IV. Another natural location for nTOF detectors is outside the NIF target bay shield wall against existing, predrilled holes. The 2-m-thick concrete walls of the target bay can be used as shielding against scattered neutrons and γ rays. There are several such locations at the NIF where detectors can be placed 17 m to 20 m from the target. The digitizing oscilloscopes can be placed nearby, thereby shortening the signal and HV cables and decreasing EMP noise pickup. The EMP noise in this location should be much less of a problem than at the target chamber wall; therefore, the minimum signal requirement in this location is 100 mV. The 20-m standoff distance is not adequate for pre-ignition and ignition targets producing 1017 to 1019 neutrons. Most of the nTOF detectors located 20 m from the target will saturate at such yields. The high neutron flux can also damage an oscilloscope and PC. Another location as far away from the target as possible with a clean flight path is needed for the ignition campaign. There is a line of sight at θ = 64° and φ = 275.62° with an opening in the target bay wall that exits outside the building just above the roof of the diagnostic building. The nTOF detectors can be installed on the roof of the diagnostic building at about 40 m from the target. OMEGA Performance Scaled to the NIF Three different types of fast detectors were used for the NIF nTOF prototype tests on OMEGA. The most-sensitive nTOF detector consists of a BC-422 scintillator coupled to a twostage MCP PMT. The PMT has a response time of about 250 ps (FWHM), a gain up to 106, and a 40-mm-diam photocathode. Detectors based on a two-stage MCP PMT are relatively sensitive to hard-x-ray and γ-ray background and are good for 203

PROTOTYPES OF NIF NEUTRON TIME-OF-FLIGHT DETECTORS TESTED ON OMEGA and DT neutron yields from 5 × 108 to 5 ×1010. Using relations derived by Lerche,12 it is estimated that, at 5 m, this detector can measure ion temperature with 15% accuracy for 109 DD neutrons at 1 keV. Better than 10% accuracy is possible at higher yields or higher ion temperatures. Al tube

34 40 mm mm

10 mm

Lead antimony

20 mm

140 mm

relatively low (109 to 1011) neutron-yield NIF implosions, in which x-ray and γ-ray background is low as well. The first prototype has a 40-mm-diam, 20-mm-thick BC-422 scintillator coupled to a Photek14 PMT240 PMT. To protect the prototype from the x-ray and γ-ray background inside the OMEGA Target Bay, it was heavily shielded on all sides by a thick lead housing (Fig. 99.72). The prototype was installed in the OMEGA Target Bay at 12.4 m from the target and connected by a 12-m-long LMR-400 cable to a 1-GHz Tektronix 684 oscilloscope. A Mini-Circuits model 15542 resistive splitter divides the detector signal between two oscilloscope channels with different sensitivity settings to increase the dynamic range of the recording system. The prototype was tested on D2 implosions on OMEGA and calibrated against the standard suite of neutron diagnostics. Figure 99.73 shows a typical scope trace of the neutron signal taken for a shot yielding 1.2 × 1011 DD neutrons and having a Ti = 4.1 keV. The measured signal was fitted by a convolution of a Gaussian and an exponential decay, as described in detail in Ref. 7. Scaled to the 5-m distance on the NIF chamber wall, this detector will have ~200 neutron interactions and produce a signal amplitude of 500 mV for a DD yield of 1 × 109 neutrons. This detector installed on the NIF target chamber wall will be sensitive to DD

Signal cable

PMT Scintillator 20 mm

HV cable

Foam 25 mm

5 mm

Lead cap

203 mm E12880

Figure 99.72 Schematic of the single-stage and two-stage MCP PMT nTOF prototypes tested on OMEGA at 12.4 m from the target.

Table 99.IV: Proposed set of nTOF detectors required for DD and DT temperature measurements for yields between 109 and 1019 neutrons.

204

N

Distance

Scintillator/ Wafer Size

Type

PMT

Yield Range

1

5m

40 mm × 20 mm

BC-422

2 MCP

1 × 109 to 5 × 1010

2

5m

40 mm × 20 mm

BC-422

1 MCP

1 × 1010 to 5 × 1011

3

5m

10 mm × 1 mm

CVD diamond

4

20 m

40 mm × 20 mm

BC-422

1 MCP

1 × 1011 to 5 × 1012

5

20 m

40 mm × 20 mm

BC-422Q

1 MCP

1 × 1012 to 1 × 1014

6

20 m

10 mm × 1 mm

CVD diamond

7

40 m

10 mm × 5 mm

BC-422Q

8

40 m

10 mm × 1 mm

CVD diamond

5 × 1014 to 1 × 1017

9

40 m

2 mm × 0.5 mm

CVD diamond

2 × 1016 to 1 × 1019

5 × 1012 to 1 × 1015

1 × 1014 to 5 × 1016 1 MCP

1 × 1014 to 1 × 1016

LLE Review, Volume 99

PROTOTYPES OF NIF NEUTRON TIME-OF-FLIGHT DETECTORS TESTED ON OMEGA was placed in a lead housing identical to that shown in Fig. 99.72 and installed in the OMEGA Target Bay at 12.4 m from the target, next to the two-stage MCP PMT prototype. This prototype was tested with DT implosions and calibrated against copper activation measurements. A scope trace of a neutron signal is shown in Fig. 99.75 (DT neutron yield of 5.0 × 1013 and Ti = 9.7 keV). If this detector is installed outside the NIF target bay at 20 m from the target, it will be sensitive to DT yields from 1012 to 1014 neutrons.

2 Fit

Signal (V)

0 –2

Signal

–4 –6 –8

560

E12882

580

600

620

640

20

Time (ns)

The second nTOF detector designed for the NIF consists of a BC-422 or BC-422Q scintillator coupled to a single-stage MCP PMT. This PMT has a slightly faster response time of about 200 ps (FWHM), a gain up to 103, and a 40-mm-diam photocathode. The single-stage MCP PMT is less sensitive (by a factor of about 103) to x-ray and γ-ray background but has less gain by the same factor. Two versions of this detector were tested on OMEGA: The first version was designed to determine the maximum sensitivity achievable using a single-stage MCP Photek PMT140 PMT, coupled to a 20-mm-thick, 40-mm-diam BC-422 scintillator. It has 14-mm lead shielding in front and a 5-mm aluminum housing on all other sides. This prototype was tested on the OMEGA chamber wall, 1.65 m from the target, with DD implosions and calibrated against the standard neutron diagnostics. Figure 99.74 shows a typical oscilloscope trace of a neutron signal from this detector (DD yield of 6.7 × 1010 neutrons and Ti = 3.2 keV). This detector installed on the NIF target chamber wall will detect ~1900 neutron interactions and produce a 500-mV signal for a DD yield of 1 × 1010 neutrons. The same detector installed outside the NIF target bay at 20 m from the target will have ~1700 neutron interactions and produce a 450-mV signal for a DD yield of 1 × 1011.

Signal –20

–40 240

LLE Review, Volume 99

280

300

320

Time (ns)

Figure 99.74 Neutron signal recorded for the singe-stage MCP PMT prototype on D2 shot 33413: yield = 6.7 × 1010 and Ti = 3.2 keV. The fit is a convolution of a Gaussian shape and a scintillator exponential decay.

5 Fit 0 Signal

–5 –10 –15 –20 220

E12883

The second version of the single-stage MCP PMT system was designed to test upper-yield limits of such detectors. This prototype had a 40-mm-diam, 20-mm-thick BC-422Q (1% benzephenone) quenched scintillator, coupled with a Photek PMT140 PMT, and operated at a gain of 5 × 102. The detector

260

E12967

Signal (V)

Figure 99.73 Neutron signal recorded for the two-stage MCP PMT prototype on D2 shot 33949: yield = 1.2 × 1011 and Ti = 4.1 keV. The fit is a convolution of a Gaussian shape and a scintillator exponential decay.

Signal (V)

Fit 0

240

260

280

300

Time (ns)

Figure 99.75 Neutron signal recorded for the single-stage MCP PMT prototype on DT shot 33797: yield = 5.0 × 1013 and Ti = 9.7 keV. The fit is a convolution of a Gaussian shape and a scintillator exponential decay.

205

PROTOTYPES OF NIF NEUTRON TIME-OF-FLIGHT DETECTORS TESTED ON OMEGA The third prototype nTOF detector tested on OMEGA used a 10-mm-diam, 1-mm-thick CVD diamond wafer. Diamond detectors have low sensitivity11 to hard-x-ray and γ-ray background and can be used at very-high DT yields. The CVD diamond detectors also have a larger dynamic range than PMTbased detectors. Such detectors11 were previously tested inside the OMEGA target chamber. This time, the CVD diamond detector was installed outside the target chamber at 2.8 m from the target and biased at 1 kV. The CVD diamond prototype was tested with DT implosions and calibrated against copper activation. A typical scope trace is shown in Fig. 99.76 (DT yield of 5.0 × 1013 and Ti = 9.7 keV). This detector installed on the NIF target chamber wall will be sensitive to DT neutrons over a yield range from 5 × 1012 to 1015. The same detector installed outside the NIF target bay at 20 m from the target will be sensitive to DT yields between 1014 and 5 × 1016 neutrons.

2 Fit

Conclusions Neutron time-of-flight (nTOF) detectors are part of the NIF core diagnostic suite providing a measurement of ion temperature and yield. Several NIF nTOF detector prototypes have been built and tested with D2 and DT implosions on OMEGA. Prototypes for low and moderate NIF neutron yields are based on fast plastic scintillators and fast photomultiplier tubes. A third prototype is based on a CVD diamond detector. A set of nTOF detectors is proposed for the NIF to measure ion temperature and DD and DT neutron yields between 109 and 1019. ACKNOWLEDGMENT

0 Signal (V)

The proposed set of detectors is a realistic, cost-effective approach to the NIF nTOF system and based on commercially available components. All of the nTOF detectors and components can be calibrated on OMEGA prior to installation and use on the NIF. The nTOF detector calibrations can be later cross checked against other yield-sensitive diagnostics, such as the PROTEX15 and activation.16

–4

This work was supported by the U.S. Department of Energy Office of Inertial Confinement Fusion under Cooperative Agreement No. DE-FC5292SF19460, the University of Rochester, and the New York State Energy Research and Development Authority. The support of DOE does not constitute an endorsement by DOE of the views expressed in this article.

–6

REFERENCES

Signal

–2

1. E. I. Moses, Fusion Sci. Technol. 44, 11 (2003).

–8 E12884

240

260

Time (ns)

Figure 99.76 Neutron signal recorded for the CVD diamond prototype on DT shot 33797: 5.0 × 1013 and Ti = 9.7 keV. The fit is a convolution of a Gaussian shape and an exponential decay.

For pre-ignition and ignition targets on the NIF, lesssensitive detectors are needed about 40 m from the target. Scaled detectors based on a single-stage MCP PMT and a CVD diamond detector are chosen for this purpose. The proposed set of nTOF detectors to measure ion temperature and DD and DT neutron yields between 109 and 1019 is summarized in Table 99.IV.

206

2. J. D. Lindl, Inertial Confinement Fusion: The Quest for Ignition and Energy Gain Using Indirect Drive (Springer-Verlag, New York, 1998). 3. T. J. Murphy, C. W. Barnes, R. R. Berggren, P. Bradley, S. E. Caldwell, R. E. Chrien, J. R. Faulkner, P. L. Gobby, N. M. Hoffman, J. L. Jimerson, K. A. Klare, C. L. Lee, J. M. Mack, G. L. Morgan, J. A. Oertel, F. J. Swenson, P. J. Walsh, R. B. Walton, R. G. Watt, M. D. Wilke, D. C. Wilson, C. S. Young, S. W. Haan, R. A. Lerche, M. J. Moran, T. W. Phillips, T. C. Sangster, R. J. Leeper, C. L. Ruiz, G. W. Cooper, L. Disdier, A. Rouyer, A. Fedotoff, V. Yu. Glebov, D. D. Meyerhofer, J. M. Soures, C. Stockl, J. A. Frenje, D. G. Hicks, C. K. Li, R. D. Petrasso, F. H. Séguin, K. Fletcher, S. Padalino, and R. K. Fisher, Rev. Sci. Instrum. 72, 773 (2001). 4. T. R. Boehly, D. L. Brown, R. S. Craxton, R. L. Keck, J. P. Knauer, J. H. Kelly, T. J. Kessler, S. A. Kumpan, S. J. Loucks, S. A. Letzring, F. J. Marshall, R. L. McCrory, S. F. B. Morse, W. Seka, J. M. Soures, and C. P. Verdon, Opt. Commun. 133, 495 (1997).

LLE Review, Volume 99

PROTOTYPES OF NIF NEUTRON TIME-OF-FLIGHT DETECTORS TESTED ON OMEGA 5. J. L. Bourgade, V. Allouche, J. Baggio, C. Bayer, F. Bonneau, C. Chollet, S. Darbon, L. Disdier, D. Gontier, M. Houry, H. P. Jacquet, J.-P. Jadaud, J. L. Leray, I. Masclet-Gobin, J. P. Negre, J. Raimbourg, B. Villette, I. Bertron, J. M. Chevalier, J. M. Favier, J. Gazave, J. C. Gomme, F. Malaise, J. P. Seaux, V. Yu. Glebov, P. Jaanimagi, C. Stoeckl, T. C. Sangster, G. Pien, R. A. Lerche, and E. Hodgson, “New Constraints for Plasma Diagnostics Development due to the Harsh Environment of MJ-Class Lasers,” to be published in Review of Scientific Instruments.

10. V. Yu. Glebov, C. Stoeckl, T. C. Sangster, and G. J. Schmid, “HardX-Ray and γ Background in Neutron Time-of-Flight Detectors on the NIF and LMJ,” to be submitted to Review of Scientific Instruments.

6. H. Brysk, Plasma Phys. 15, 611 (1973).

13. T. J. Murphy et al., Rev. Sci. Instrum. 72, 850 (2001).

7. T. J. Murphy, R. E. Chrien, and K. A. Klare, Rev. Sci. Instrum. 68, 610 (1997).

14. Photek Ltd., St. Leonards-on-Sea, United Kingdom.

8. Saint-Gobain Crystals, Newbury, OH 44065. 9. R. A. Lerche, D. W. Phillion, and G. L. Tietbohl, in Ultrahigh- and High-Speed Photography, Videography, and Photonics ’93, edited by P. W. Roehrenbeck (SPIE, Bellingham, WA, 1993), Vol. 2002, pp. 153–161.

LLE Review, Volume 99

11. G. J. Schmid, R. L. Griffith, N. Izumi, J. A. Koch, R. A. Lerche, M. J. Moran, T. W. Phillips, R. E. Turner, V. Yu. Glebov, T. C. Sangster, and C. Stoeckl, Rev. Sci. Instrum. 74, 1828 (2003). 12. R. A. Lerche and B. A. Remington, Rev. Sci. Instrum. 61, 3131 (1990).

15. M. J. Moran, V. Yu. Glebov, R. Rygg, and B.-E. Schwartz, “PROTEX: A Proton-Recoil Detector for ICF Fusion Neutrons,” submitted to Review of Scientific Instruments. 16. G. W. Cooper and C. L. Ruiz, Rev. Sci. Instrum. 72, 814 (2001).

207

Loading...

Prototypes of NIF Neutron Time-of-Flight Detectors Tested on OMEGA

PROTOTYPES OF NIF NEUTRON TIME-OF-FLIGHT DETECTORS TESTED ON OMEGA Prototypes of NIF Neutron Time-of-Flight Detectors Tested on OMEGA Introduction Th...

121KB Sizes 1 Downloads 7 Views

Recommend Documents

#nif hashtag on Twitter
Embed Tweet. 【拡散希望】 【提供】 12/31 西日本アイドルフェスティバル BO〜BP列2連番orCG〜CK

Omega and NIF Target Campaigns at LANL - Los Alamos National
Aug 23, 2012 - Target design requires a 20um thick PVD coating of aluminum/titanium onto one half of the split foam. Spe

Advantages of Neutron - Hindawi
Feb 3, 1989 - Neutron diffraction texture analysis is based on pole figure measurement followed by pole figureinversion.

prototypes - yayasan inovasi malaysia
2. SBP INTEGRASI GOPENG, GOPENG. 3. SEK. RAJA PEREMPUAN TAAYAH, IPOH. State: Perak. Parficipafing Schools: 1. SMK TTDI J

Metal Detectors
Features and circuitry for hunting saltwater beaches; tunes out highly mineralized black sand to detect ..... A PINPOINT

An Introduction to Neutron Scattering - Neutron Sciences
Magnetic scattering involves a form factor – FT of electron spatial distribution. • Electrons are distributed in spa

OMEGA ICF experiments and preparation for direct drive ignition on NIF
(NIF) ignition target based on an 'all-DT' design: a spherical target of ∼3.4 mm diameter, with a. 1–2 µm CH wall t

replica omega seamaster hinta Omega Speedmaster , Opinion on the
Results 1 - 10 of 29 - replica omega seamaster hinta I would really like to hear peoples opinions on the best model spee

Residential Treatment - Prototypes
Residential Treatment. Prototypes' residential treatment programs are designed to help women remain successful in their

NIF A2, NIF A3 - DOCUMENTS.TIPS
Jul 7, 2015 - NIF A-2 El objetivo de la NIF A-2 es describir un eslabón más en la jerarquía de conceptos plasmada en