Two

CHAPTER 2

Lung Volumes

 Introduction

69

 Lung Subdivisions: Volumes and Capacities  Measurement of VC

70

73

 Measurement of FRC

77

 Measurement of Lung Volumes by Other Methods  Reference Values

100

 Infection Control

101

 Case Presentations

99

102

 Self-Assessment Questions 110

Introduction In the pulmonary function laboratory, the measurement of lung volumes or measurement of static lung volumes usually refers to the measurement of total lung capacity (TLC), residual volume (RV), functional residual capacity (FRC), and vital capacity (VC). These measurements are essential for analyzing lung function; they provide information to further the diagnostic process and assess therapy. There are two major steps in measuring lung volumes: (a) determining FRC and (b) measurement of slow vital capacity (SVC) and its subdivisions. FRC is most commonly determined with one of three basic techniques: (a) body plethysmography, (b) multiple-breath closed-circuit helium (He) dilution, or (c) multiple-breath opencircuit nitrogen (N2) washout. This chapter will discuss these techniques in detail focusing on instrumentation, relevant physiology, calculations, and testing techniques. In addition, this chapter will discuss the measurement of SVC and its subdivisions. Other methods of measuring static lung volumes include the single-breath N2 test, the single-breath He (or other inert gas) test, the chest roentgenogram (x-ray), and computed tomography (CT). The single-breath N2 and chest x-ray methods were much less commonly used as of 2010 and thus will be discussed only briefly. The single-breath He (or other inert gas) method

69

c02.indd 69

5/26/2011 6:29:24 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

70

CHAPTER 2

Lung Volumes

is performed in conjunction with the single-breath carbon monoxide diffusing capacity test (DL,CO) and is discussed only briefly in this chapter. It is discussed more extensively in Chapter 3. The use of CT scans has increased significantly and will be briefly discussed.

Lung Subdivisions: Volumes and Capacities The total lung volume can be divided into several subdivisions or compartments (Figure 2.1), which can then be grouped into volumes and capacities. There are four volumes: inspiratory reserve volume (IRV), tidal volume (TV), expiratory reserve volume (ERV), and residual volume (RV). Two or more of these volumes make up a capacity. For example, the sum of IRV and TV is inspiratory capacity (IC). There are four capacities: vital capacity (VC), inspiratory capacity (IC), functional residual capacity (FRC), and total lung capacity (TLC). TV (Figure 2.2A) is the volume of air inspired and expired with each breath during normal breathing. The end of the inspiratory phase is called the end-inspiratory level, and the end of the expiratory phase is called the end-expiratory level. IRV (Figure 2.2B) is the maximum volume of air that can be inhaled from the endinspiratory level during quiet or normal tidal breathing. The ERV (Figure 2.2C) is the maximum

Figure 2.1 Lung volume compartments and subdivisions based on a volume–time spirogram. The four volumes are IRV, TV, ERV, and RV. The capacities (VC, IC, FRC, and TLC) consist of two or more volumes. Source: Adapted from Forster RE, DuBois AB, Briscoe WA, Fisher AB. The lung: Physiologic basis of pulmonary function tests. 3rd ed. Chicago: Year Book Medical Publishers, 1986. Maximal inspiratory level

IRV IC

Resting inspiratory level

TLC

VC TV

Resting expiratory level ERV FRC

Maximal expiratory level RV

c02.indd 70

RV

5/26/2011 6:29:25 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Lung Subdivisions: Volumes and Capacities

71

volume of air that can be exhaled from the end-expiratory level during quiet or normal tidal breathing (i.e., from FRC). RV is the volume of air remaining in the lungs at the end of a maximum expiration. Unlike the other three volumes, RV must be measured indirectly using a two-step process. First, the

Figure 2.2 The three volumes that can be measured directly from the spirogram. A. Tidal volume (TV). B. Inspiratory reserve volume (IRV) C. Expiratory reserve volume (ERV). A

Volume

TV

Time B IRV

Volume

TV

Time C

Volume

TV ERV

Time

c02.indd 71

5/26/2011 6:29:25 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

72

CHAPTER 2

Lung Volumes

FRC is measured using one of several techniques. Then an SVC is obtained and subdivided (i.e., TV, IRV, ERV). When the FRC and SVC have been obtained, RV can be calculated in the following way: RV  FRC  ERV The VC is the volume change at the mouth and can be an expiratory VC or an inspiratory VC. The expiratory VC is the volume of air that can be exhaled from the lungs after a maximum inspiration. The inspiratory VC is the volume of air that can be inhaled from a position of full expiration to full inspiration. The VC can also be performed forcefully or slowly. When done forcefully it is called the forced vital capacity (FVC), which is defined as the volume of air that is exhaled forcefully starting from a position of full inspiration and ending at complete expiration. When the VC is exhaled slowly, it is called the SVC, which is defined as the volume of air that is exhaled without force from a position of full inspiration to full expiration. The VC can also be described as the sum of the TV, IRV, and ERV. In a healthy individual, the VC makes up approximately 70% of the TLC. The IC is the maximum amount of air that can be inhaled from the TV end-expiratory level. It is the sum of the TV and IRV. This capacity usually makes up 60% to 70% of the VC in healthy individuals. The FRC is the volume of air remaining in the lungs at the TV end-expiratory level. At this point in the respiratory cycle (i.e., end expiration), the elastic force of the chest wall (acting to expand the chest) is exactly balanced by the elastic force of the lungs (acting to deflate the lungs). As shown in Figure 2.1, the FRC consists of the ERV and RV, and its determination is a critical step in measuring lung volumes. The TLC is the volume of air in the lungs after a maximum inspiration. It consists of all four volumes (IRV, TV, ERV, and RV) and two capacities (IC and FRC). The size of the lung volume compartments varies among individuals, depending on age, gender, height, weight, race, and disease. In healthy individuals the lung volume compartments vary from the factors previously described, but the proportional relationships are similar. Typically, in healthy individuals the FRC is approximately 40% to 50% of the TLC, and the RV is approximately 25% to 30% of the TLC (Figure 2.3). In individuals with lung disease, forced spirometry alone does not always provide adequate information. For example, in individuals with airflow obstruction (e.g., emphysema), it is common to see a reduced FVC and forced expiratory volume in 1 second (FEV1) and a reduced FEV1/FVC ratio. Similarly, an individual with a restrictive process (e.g., interstitial lung disease) will also have a reduced FVC and FEV1; however, the FEV1/FVC ratio is typically increased. Although the FEV1/FVC ratio may be helpful in differentiating obstructive and restrictive diseases, it is not the acid test. The only true way to differentiate the two disease processes is to measure all the lung volume compartments. In addition to a reduced FVC, FEV1, and FEV1/FVC ratio, patients with airflow obstruction usually have an increased FRC and a decreased ERV. Hence, the RV and the TLC are increased (Figure 2.3). The amount of increase depends on the severity of obstruction. As noted earlier, in individuals with restrictive patterns, the FVC and FEV1 are reduced and the FEV1/FVC ratio is usually increased. Additionally, the lung volume compartments are all proportionally reduced (Figure 2.3).

c02.indd 72

5/26/2011 6:29:25 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of VC

73

Figure 2.3 SVCs in healthy and diseased patterns plotted as a percentage of total lung volume. The healthy pattern shows maximum inspiration (TLC) at 100% of total lung volume, and the maximum expiration (RV) is approximately 20% of total lung volume. The airflow obstruction pattern shows a reduced SVC with elevated TLC and RV. The restrictive pattern also shows a reduced SVC, as well as a reduced TLC and RV. 120

100

80

% Total lung volume

60

40

20

0

Healthy

Airflow obstruction

Restriction

In mixed obstructive and restrictive disorders, some compartments are normal, some are increased, and some are reduced, making interpretation difficult. Table 2.1 shows the various static lung volume compartments and their status for a particular disease pattern.

Measurement of VC The SVC is the maximum volume of air that can be slowly exhaled from the lungs after a maximum inspiration to a full expiration. In healthy individuals, there is little difference between the SVC and the FVC. However, in individuals with airflow obstruction, the FVC maneuver causes gas trapping and thus is smaller than the SVC.

c02.indd 73

5/26/2011 6:29:25 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

74

CHAPTER 2

Lung Volumes

Table 2.1 Lung Volume Compartments and Their Status for Obstructive, Restrictive, and Mixed Obstructive and Restrictive Lung Disease Patterns

VC

Obstructive

Restrictive

Mixed

D or N

D

D

TLC

I

D

N or D

IC

N

D

N or D

FRC

I

D

N or D

ERV

D or N

D

D

I

D

N or D

RV

I, increased; D, decreased; N, normal.

The same instrumentation is used to measure SVC and FVC. Many computerized pulmonary function testing systems contain software that can measure the SVC and its subdivisions. It is recommended that if both SVC and FVC are to be measured, SVC should be performed before FVC maneuvers because of the potential for muscular fatigue and gas trapping.1 The SVC can be done as an inspiratory SVC (often referred to as IVC) or as an expiratory SVC (often referred to as EVC). As in forced spirometry and prior to testing, provide the patient with instructions on the appropriate technique, and demonstrate the SVC maneuver. It is important to stress that the maneuver is not forced but is done in a relaxed manner except near end-inspiration and endexpiration. A nose clip should be worn by the patient during the maneuver. The maneuver for performing the expiratory SVC (or EVC) should include several tidal breaths to establish a stable FRC level, followed by a maximum inspiration and then a slow, complete expiration (Figure 2.4). The maneuver for performing the inspiratory SVC (or IVC) includes several tidal breaths to establish a stable FRC level, followed by a maximum expiration and then a slow, complete inspiration (Figure 2.5). One additional variation of measuring the SVC that is less commonly used employs the concept that the VC is the sum of the TV, IRV, and ERV. After breathing at TV for several breaths, the patient inspires maximally and then returns to quiet breathing at the tidal breathing level. After again breathing at TV for several breaths, the patient slowly expires maximally. This two-step technique (Figure 2.6) measures the SVC in parts, which then can be summed.

Repeatability and Reporting As with forced spirometry, at least three acceptable VC maneuvers should be obtained. The two highest SVC values should agree within 0.150 liters. If this is not achieved, additional maneuvers should be performed.1 The largest value from at least three acceptable maneuvers should be reported. For IC, the mean of at least three acceptable maneuvers should be reported.1

c02.indd 74

5/26/2011 6:29:26 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of VC

75

Figure 2.4 Typical SVC maneuver as expressed on a volume–time spirogram with tidal breathing followed by a maximum inspiration and then a slow, complete expiration.

Inspiration Expiration Volume Tidal breathing

Time

Figure 2.5 The reverse of the typical expiratory SVC is the inspiratory SVC. On a volume–time spirogram, tidal breathing is followed by a slow, complete expiration and then a maximum inspiration.

Inspiration

Volume Tidal breathing Expiration

Time

c02.indd 75

5/26/2011 6:29:26 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

76

CHAPTER 2

Lung Volumes

Figure 2.6 The two-step VC, as shown on a volume–time spirogram, consists of tidal breathing followed by a maximal inspiration, then a return to tidal breathing followed by a complete expiration.

Inspiration Expiration

Volume Tidal breathing Expiration

Time

Determining Lung Subdivisions The determination of RV and TLC can be complicated. There are two recommended methods to determine these values after FRC has been determined.2 The first and preferred method is to measure ERV immediately after the acquisition of the FRC measurement followed by slow inspiratory vital capacity. All measurements should be performed as a linked maneuver, that is, without the patient coming off the mouthpiece prior to completion of the maneuvers.2 The reported values for RV and TLC are: RV  Mean of FRC  Mean of ERV TLC  Reported value for RV  Largest IVC The second recommended method includes the measurement of IC immediately after the FRC determination.2 This method may be necessary in patients with severe dyspnea who cannot perform the ERV immediately after the FRC measurement. With this approach, patients can come off the mouthpiece between successive linked FRC and IC determinations and also between the separate VC maneuvers. The reported values for RV and TLC using this second approach are: RV  Mean TLC  Largest VC measured TLC  Mean of the three largest sums of FRC values and linked IC values

c02.indd 76

5/26/2011 6:29:26 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of FRC

77

Measurement of FRC As noted previously, determining FRC is one of two main steps in determining lung volumes. The three most commonly used FRC determination techniques are: (a) body plethysmography (also called body box and abbreviated FRCpleth), (b) multiple-breath He dilution (abbreviated FRCHe) and (c) multiple-breath N2 washout (abbreviated FRCN2). In healthy subjects and individuals with pure restrictive lung disease, these three methods show good agreement.3-6 In individuals with obstructive lung disease, however, there is generally poor agreement between gas dilution values (He dilution and N2 washout) and values from the body box. The body plethysmograph or body box measures the volume of gas in the thoracic cage and is considered the most accurate method of measuring FRC. However, the body box is more expensive, technically more complicated, and requires more patient cooperation and effort than gas dilution–washout methods. Additionally, the body box may overestimate FRC slightly.7-11 The suspected overestimation error stems from the fact that in severe obstructive lung disease, alveolar pressure may be underestimated. This matter is discussed later in this chapter. The multiple-breath gas dilution–washout methods are relatively simple to perform and require little effort from the patient. However, in individuals with obstructive lung disease, the gas dilution–washout methods underestimate the true FRC because they measure only those areas of the lung in communication with the mouth. They do not measure areas of the lung that contain trapped gas. A large number of obstructive lung disease patterns are thus falsely classified as mixed obstructive–restrictive lung disease patterns because FRC is underestimated. Both the gas dilution–washout and plethysmographic methods for determining FRC are considered acceptable.2,3,12 The gas dilution methods underestimate the true FRC in patients with airflow obstruction. The plethysmograph method includes those air spaces not measured by gas dilution–washout methods and thus is the method of choice in patients with airflow obstruction. The use of both methods provides useful information about the volume of noncommunicating gas (i.e., trapped gas).

Measuring FRC with Body Plethysmography The body plethysmograph or body box is the first of three methods that will be discussed to measure FRC. According to Comroe,13 Pfluger was the first to apply Boyle’s law to measure RV in 1882 by constructing a large metal container. The English translation of the German name for Pfluger’s device was man-box or man-can. However, DuBois and coworkers are usually credited with the modern application.6 To measure FRCpleth, the patient, sitting inside a sealed box, breathes quietly for several breaths and then pants against a closed shutter. The gas volume trapped in the lungs when the shutter is closed can then be measured by applying Boyle’s law. As previously mentioned, this method is generally considered the most accurate of the three methods for measuring FRC because it measures the total gas volume in the thoracic cage or thoracic gas volume (TGV). The body box is also used to measure airway resistance (Raw). The patient uses the same panting technique, or even a quiet breathing technique in some systems, but first with the shutter open and then with the shutter closed. Raw is used to assess airflow obstruction rather than lung volumes. It is discussed in Chapter 4.

c02.indd 77

5/26/2011 6:29:26 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

78

CHAPTER 2

Lung Volumes

The discussion of FRCpleth determination includes: (a) physiology and instrumentation and (b) testing technique. Physiology and Instrumentation The operating principle of the body box is based on Boyle’s law, which states that a volume of gas at constant temperature varies inversely with the pressure applied to it. In other words, when gas in a closed container is compressed, the volume decreases while the pressure inside the container increases. The reverse is true if the gas in the closed container is decompressed. Mathematically, Boyle’s law is written as: P1V1  P2V2

(Eq. 2.1)

The individual to be tested sits in the sealed body box, attaches to the mouthpiece, wears a nose clip, and breathes quietly. When a valve (i.e., shutter) to which the mouthpiece is connected is closed at the TV end-expiratory position (i.e., FRC), it traps that volume of gas in the lungs. By trying to pant in and out against the closed shutter, the patient compresses and decompresses that trapped volume of gas in the thorax while the temperature remains constant. By panting out (i.e., against the shutter), the chest wall moves inward, which compresses the thoracic gas. Because the chest wall moves inward there is a proportional decompression of the gas in the sealed body box. Conversely, the decompression of the thoracic gas results in a proportional compression of the gas in the box. Because there is no airflow, the pressure changes inside the lung (measured at the mouth) and volume changes in the TGV (measured by changes in body box volume) allow the trapped TGV to be determined by applying Boyle’s law. The derivation of the formula for determining TGV using Boyle’s law is shown in Appendix C. The final form of the formula as shown in Appendix C is: V1 

P1V P

(Eq. 2.2)

where P1  Alveolar pressure V  Change in body box volume when panting P  Change in alveolar pressure (measured at the mouth) when panting against a closed shutter V1  TGV when the shutter is closed, usually at FRC In practice, the operator observes a computer monitor that displays the relationship between alveolar pressure measured at the mouth and body box volume or pressure (Figure 2.7). The V and P are measured as the slope of the line fitting the panting lines. The slope (rise over run) is P/V, but equation 2.2 requires V/P. Thus, the inverse of the slope is required. The inverse of the slope is then multiplied by the calibration factors for body box volume and mouth pressure. Thus, equation 2.2 can be rewritten as:

c02.indd 78

5/26/2011 6:29:26 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of FRC

TGV = P 

79

Body box calibration factor 1  Slope Mouth pressure calibration factor

where TGV  Thoracic gas volume P  Barometric pressure in cm H2O less water vapor Slope  P/V Let us use some hypothetical numbers. The closed shutter panting produced the display shown in Figure 2.7. Angle x is measured to be 45 degrees, and the slope or tangent of 45 degrees is 1.00. The body box pressure calibration factor, which is typically determined by injecting a known volume into the sealed body box and measuring the horizontal deflection, is 10 mL/cm. The mouth pressure calibration factor, which is typically determined by applying a known pressure to the mouth pressure transducer and measuring the vertical deflection,

Figure 2.7 Plethysmograph display showing mouth pressure and body box pressure relationship during closed-shutter panting. The dashed line is the fit of the several panting lines. Angle x is used to calculate TGV. Fit line 10

5 Angle x Mouth pressure

0

−5

−10

0 Body box pressure

c02.indd 79

5/26/2011 6:29:27 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

80

CHAPTER 2

Lung Volumes

is 2.5 cm H2O/cm. If the barometric pressure is 760 mmHg and the water vapor pressure at 37 C is 47 mmHg, then P  760  47  713 mmHg  970 cm H2O* Plugging these values into the rewritten equation 2.2 results in the following: TGV = P 

Body box calibration factor 1  Slope Mouth pressure calibration factor

TGV = 970 cm H 2 O 

10 mL/cm 1  1 2.5 cm H 2 O/cm

TGV  970 cm H2O  4 mL/cm H2O TGV  3,880 mL or 3.880 liters The alveolar pressure changes caused by the compression and decompression of air in the lungs are estimated at the mouth. As mentioned earlier, the assumption that alveolar pressure equilibrates and is correctly measured at the mouth when the glottis is open and the cheeks are held firmly has been questioned.7-10 In the presence of severe airflow obstruction there is a phase lag between pressures at the mouth and in the alveoli. This leads to underestimation of alveolar pressure and overestimation of TGV. This problem can be minimized by reducing the patient’s panting frequency to 0.5 to 1.0 breaths/sec. The TGV changes caused by the compression and decompression of the chest wall during the panting maneuver are estimated by measuring changes in the body box volume. The sealed box allows for measurement of the small changes in volume (e.g., 200 to 500 mL). The three major types of body boxes (variable-pressure box, flow box, and volume-displacement box) are classified based on how each type measures the body box volume change. Variable-Pressure Box The variable-pressure box (Figure 2.8) uses a pressure transducer to measure body box pressure changes that are caused by the compression and decompression of the chest in a constant volume container. Hence, it is also known as a constant-volume box. This transducer, which is attached to the wall of the box, is calibrated by injecting known volumes into the sealed box, creating a relationship between box volume change and box pressure change. This type of box requires a correction for body weight and frequent venting of excess pressure caused by rising temperatures when the patient is inside. Flow Box The flow box (Figure 2.9) measures box volume changes within a constant pressure chamber using a large pneumotachograph (which measures flow) placed in the box wall. When the thorax compresses and decompresses, the air in the body box will flow in and out of the box *1.36 converts mmHg to cm H2O.

c02.indd 80

5/26/2011 6:29:27 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of FRC

81

Figure 2.8 Variable-pressure body plethysmograph (constant volume). The patient, wearing a nose clip, attaches to the mouthpiece and breathes through a shutter–pneumotach apparatus. The shutter (S) is open for tidal breathing and for measurements of Raw and closed for measurements of TGV. When the shutter is closed, mouth pressure is measured by a transducer (T2). The pneumotach (P) measures flow using the transducer T1, and in modern body plethysmographs the flow signal is electronically integrated to obtain volume. The body plethysmograph (or body box) pressure is measured by a transducer (T3). The signals from the three transducers are processed by a computer. Excess body box pressure from temperature changes caused by the patient sitting in the closed box is vented through a valve (V).

T2

S

P T1

T3

V

Pressure box

through this pneumotachograph. The measured flow is then integrated to obtain volume. The frequency response can be corrected. The flow and mouth pressure signals must be in phase, either by mechanical adjustments or computer software. The flow box can be converted into a pressure box simply by occluding the pathway through the pneumotachograph, which combines the flow and pressure box approaches. This type of body box is commonly sold commercially. This combination allows for the measurement of TGV using the pressure box mode (the pathway to the pneumotachograph is closed) and the flow–volume curve using the flow box mode (the pathway to the pneumotachograph is open).

c02.indd 81

5/26/2011 6:29:27 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

82

CHAPTER 2

Lung Volumes

Figure 2.9 Flow body plethysmograph (variable volume). The patient, wearing a nose clip, attaches to the mouthpiece and breathes through a shutter–pneumotach apparatus. The shutter (S) is open for tidal breathing, measurement of Raw, and spirometry. It is closed for measurement of TGV. When the shutter is closed, mouth pressure is measured by a transducer (T2). The pneumotach (P1) measures flow through transducer T1. The flow can then be electronically integrated to obtain volume. Modern systems allow the user to perform SVC and FVC. Changes in plethysmograph (body box) volume, which occur with chest-wall movement, are measured by electronic integration of flow through pneumotach P2 and transducer T3. The signals from the three transducers are processed by a computer.

T2

S

P1

T1

P2

T3

Flow box

Volume-Displacement Box The volume-displacement box (Figure 2.10) measures box volume changes within a constant pressure chamber with an attached spirometer. It is easily calibrated by injecting a known volume into the sealed box. The frequency response of the spirometer must be taken into account or corrected, and the warming of the air in the plethysmograph can be overcome by air conditioning. The volume-displacement box can measure large volume changes (e.g., VC) and both the volume expired at the mouth and the volume compressed by the chest wall. There

c02.indd 82

5/26/2011 6:29:27 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of FRC

83

can be a considerable difference between these two volumes in individuals with severe airway obstruction.14 In the past, the relationship between mouth pressure (which estimates alveolar pressure) and box volume changes was hand measured. Today, excellent software can calculate this measurement and make the body box easier to use. Additionally, software and instrumentation advances for the body box allow for the performance and measurement of many pulmonary function tests, including spirometry (SVC and FVC), diffusing capacity, single- and multiple-breath N2 washout, pressure–volume characteristics, and maximal inspiratory and expiratory pressures.

Figure 2.10 Volume-displacement body plethysmograph (variable volume). The patient, wearing a nose clip, attaches to the mouthpiece and breathes through a shutter–pneumotach apparatus, which is usually located outside the plethysmograph (body box). The shutter (S) is open for tidal breathing, measurement of Raw, and spirometry. It is closed for measurement of TGV. When the shutter is closed, mouth pressure is measured by a transducer (T2). The pneumotach (P) measures flow through transducer T1. The flow can then be electronically integrated to obtain volume. Changes in body box volume, which occur with chest-wall movement, are measured by a volume-displacement spirometer (VS) and a linear volume-displacement transducer (LVDT). The type of spirometer shown is a Krogh water-sealed spirometer. Signal processing is done by a computer, and modern systems allow the user to perform SVC and FVC. Volume box

T2

S

T1

P

VS LVDT

AC

c02.indd 83

5/26/2011 6:29:28 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

84

CHAPTER 2

Lung Volumes

Testing Technique The testing technique for measuring FRCpleth includes: (a) equipment preparation and calibration, (b) patient preparation and instructions, (c) testing, and (d ) quality control. Equipment Preparation and Calibration Calibrate the body box every day it is used. Follow the calibration process described in the operator’s manual provided by the manufacturer. Usually the process consists of calibrating the box volume with a calibration syringe and the mouth pressure transducer with a water manometer. Additionally, calibrate any flow-sensing device attached to the breathing tube with a calibration syringe. Maintain a log of calibration results showing date and time, calibration values, barometric pressure, and the identification of the individual performing the calibration. It is common to print the calibration pages and store them in a notebook for review. Patient Preparation and Instructions Give the patient some instructions before he or she enters the box, such as how to use the mouthpiece and nose clip. Tell the patient that after he or she sits inside the door will be closed. Many patients are apprehensive about being closed in a small space. Assure the patient that the door can be opened at any time between maneuvers. Additionally, most boxes today are constructed with clear plastic walls, which promote a feeling of openness. Patients who are receiving intravenous (IV) or O2 therapy present a problem. If the patient is receiving IV therapy, the bags and pumps must be temporarily disconnected before testing in the body box. Similarly, patients receiving O2 therapy (including transtracheal) should have their O2 temporarily turned off while measurements are made. Carefully explain the panting maneuver. Usually the patient is asked to pant (shallow breaths) at a rate of 0.5 to 1 breath/sec. When the shutter closes and the airflow stops, panting is more difficult. Explain that although air movement really stops when the shutter closes, the pressure of the panting is being measured, and therefore the patient must try to continue the panting maneuver. Next, demonstrate the open (if Raw is to be measured) and closed-shutter maneuvers. Practicing with the patient is the next step. With the door closed and sealed, communication is hampered. Most of the intercom systems on the boxes do not provide good sound quality and therefore can cause delays as the patient tries to understand and perform this tricky maneuver. One recommendation is to leave the door open and show the patient what it feels like to pant against the closed shutter and give additional instructions or suggestions. When you think the patient is ready, seal the door and proceed with the testing. The patient still may have some problems with the maneuver (described next), but you can do any fine-tuning over the intercom system. Testing The common problems encountered in getting the patient to do the panting maneuver correctly include incorrect panting frequency (i.e., too fast or too slow), failure to inspire and expire against the closed shutter, panting too hard, lips failing to seal around mouthpiece, and glottis closure. Additionally, unsatisfactory results may occur when the patient allows the cheeks to puff in and out with the closed-shutter panting. Correct this by having the patient place his or her hands on the cheeks and prevent them from moving (Figure 2.11).

c02.indd 84

5/26/2011 6:29:28 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of FRC

85

Figure 2.11 To keep a patient’s cheeks from puffing in and out during the closed-shutter maneuver, his or her hands should be placed on the cheeks and pressed firmly toward the face.

Careful instruction and good feedback can reduce problems. During the closed-shutter panting maneuver, carefully monitor the graphic plot showing the change in mouth pressure and box volume. Ideally, you will observe a series of superimposed lines (Figure 2.12). Looping or bending can occur with glottis closure or box leaks or when the cheeks move (Figure 2.13). As noted earlier, the relationship (i.e., the slope) of mouth pressure and body box volume changes during the closed-shutter maneuver was hand measured in the past. Current body box systems contain software that can estimate this slope. The advantage of allowing the computer to analyze all the data points and fit a line include: (a) faster analysis, (b) analysis of each individual panting breath, (c) greater accuracy by elimination of observer bias, and (d ) better consistency.15 It has been recommended that the operator acquire at least three to five satisfactory panting maneuvers, with at least three FRCpleth values that agree within 5% (i.e., the difference between the highest and lowest value divided by the mean is less than 0.05).2 If significant variability exists, obtain additional values. The mean value of all acceptable and repeatable FRCpleth values should be reported in liters at BTPS, rounded to two decimals (e.g., 3.13 liters).

c02.indd 85

5/26/2011 6:29:28 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

86

CHAPTER 2

Lung Volumes

Figure 2.12 Plethysmograph display of a properly done closed-shutter panting maneuver. It demonstrates the relationship between mouth pressure and body box pressure as a series of overlapping straight lines.

10

5

Mouth pressure

0

−5

−10

0 Body box pressure

Figure 2.13 Plethysmograph displays of poorly done panting maneuvers. The relationship between mouth pressure and body box pressure is a series of open (A) or bent loops (B), usually caused by thermal drifting or patient technique (panting too rapidly or panting too deeply), which makes measurement difficult. The dashed lines represent the best fit if these measurements were used, but the operator should work hard with the patient to correct the problems and achieve a better measurement.

c02.indd 86

5/26/2011 6:29:29 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of FRC

87

A

10

5

Mouth pressure

0

−5

−10

0 Body box pressure B

10

5

Mouth pressure

0

−5

−10

0 Body box pressure

c02.indd 87

5/26/2011 6:29:29 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

88

CHAPTER 2

Lung Volumes

Quality Control The accuracy of the body plethysmograph in measuring FRC using a mechanical model has been described.16,17 This technique uses a container of known volume as a model lung (e.g., 3- or 4-liter glass flask filled with copper wool, with a two-holed rubber stopper at the opening). One hole is connected to the body box mouthpiece, and the second hole is connected to a rubber bulb. The bulb can be squeezed at the appropriate frequency by an individual sitting inside the box (holding his or her breath). The volume measured should equal the volume in the flask less the volume of copper wool (calculated from weight and density) within 50 mL or 3%, whichever is greater based on a mean of five determinations.18 Some manufacturers of body plethysmographs provide variations of this early mechanical model. Biological controls (healthy nonsmoking individuals) should be tested at least monthly or whenever errors are suspected. Measurements should include FRCpleth, RV, and TLC. Values that differ by more than 10% for FRCpleth and TLC, or more than 20% for RV from previously established means on the same biological control, suggest errors. Another approach to ensure accuracy is to measure FRCpleth and compare the value to FRCHe and/or FRCN2. Values for FRC using the different techniques should be very similar in healthy individuals.

Measuring FRC by Multiple-Breath Closed-Circuit He Dilution The second method for determining FRC is by multiple-breath closed-circuit He dilution (FRCHe).19,20 This method involves diluting He, an inert gas, in the lungs by rebreathing the gas in a closed system over a short time (usually 2 to 10 minutes). This method is widely used in pulmonary function laboratories because: (a) it is a very easy test for the patient to perform, since it requires only tidal breathing and minimal learning and effort, and (b) the instrumentation is simple and inexpensive. However, as mentioned earlier, a major drawback of this method is that it measures only the lung volume in communication with the mouth. This becomes a problem in individuals with airflow obstruction because significant amounts of lung volume may not communicate with the mouth. The result underestimates FRC, which leads to an underestimation of RV and TLC. The discussion of this method includes: (a) instrumentation, (b) physiology and calculations, and (c) testing technique. Instrumentation The most commonly used He analyzer operates on a thermal conductivity principle. This principle is based on the facts that gases are able to conduct heat and different gases conduct heat at different rates. By introducing gas molecules into a sampling chamber containing a heated wire, the temperature of that wire decreases, allowing more current to flow through the wire. To be specific for He, other gases must either be removed or be compared with a reference. Therefore, CO2 is scrubbed from the sample gas using a chemical absorber, and O2 and N2 are negated by comparing the thermal conductivity of the sample gas with a dry room-air reference. A Wheatstone bridge is incorporated to measure the resistance difference of the sample and reference chambers. Moisture in the circuit gas results in changes in H2O vapor pressure, which affects the He analyzer. Thus, H2O is also removed with a chemical absorber.

c02.indd 88

5/26/2011 6:29:29 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of FRC

89

A small pump draws the gas from the breathing circuit and passes it over the chemical absorbers (one for CO2 and one for H2O), passes it to the He analyzer, and then passes it back into the breathing circuit. A volume-displacement spirometer with a capacity of at least 7 liters is commonly used. A mixing fan is incorporated to circulate and mix the air throughout the breathing circuit. A breathing valve, corrugated tubing, and mouthpiece are used and become part of the circuit dead space. These items are usually disassembled and disinfected or sterilized between uses. The complete system is shown in Figure 2.14.

Figure 2.14 Closed-circuit He dilution for determining FRC. A. The known concentration and volume of He in the spirometer system is separated from the patient by a closed valve. B. The valve is opened, and He (dots) is redistributed by rebreathing until it is equilibrated in the lungs and the circuit.

A Valve closed to patient

CO2 absorber

Helium Analyzer

B Valve open to patient

CO2 absorber

Helium Analyzer

c02.indd 89

5/26/2011 6:29:30 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

90

CHAPTER 2

Lung Volumes

Physiology and Calculations The closed-circuit He dilution technique uses a spirometer and breathing circuit that contains a known volume (VS) and concentration (CS) of He (Figure 2.14). The patient, wearing a nose clip, is connected to the mouthpiece. After a valve is opened (connecting the patient to the spirometer and test gas, which contains He, O2, and N2), the patient breathes into and out of this closed spirometer circuit. The initial concentration of He in the spirometer (CS), commonly 10%, is diluted proportionately by adding the patient’s lung volume (VL), resulting in a final He concentration that reflects the concentration in both the spirometer circuit and the lungs (CLS). If the patient is turned in to the spirometer at TV end-expiratory level (FRC), then the VL that will be measured is FRC. However, the patient could be turned in above or below FRC, allowing a lung volume other than FRC to be measured. If this occurs, it is referred to as a switching error, and a mathematical correction is usually made to ensure that FRC is correctly measured (see more discussion on switching error in the next section). When the patient is turned in to the spirometer circuit and begins to breathe and rebreathe the circuit gas, the rebreathing will continue until the He concentration is equilibrated (Figure 2.14). In healthy individuals, the equilibration time takes approximately 2 to 3 minutes. However, in individuals with obstructive lung disease, it may take as long as 5 to 10 minutes, but it rarely exceeds 10 minutes. One technique that is sometimes used to speed up the equilibration process and thus reduce testing time is a periodic deep breath. Looking at the calculation more closely, it can be seen that: Volumes and concentrations at beginning of test



Volumes and concentrations at end of test

or VSCS  VLCL  VSCLS  VLCLS Because CL  0 at the beginning of the test: VSCS  VSCLS  VLCLS Solving for VL: VL 

VS CS  CLS VS CLS

VL 

VS (CS  CLS ) CLS

If the concentration of He in the spirometer at the start of the test (CS) is called C1, and the concentration of He in the spirometer and patient’s lungs at the end of the test (CLS) is called C2, and VL is called FRC, then:

c02.indd 90

5/26/2011 6:29:31 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of FRC

FRC 

VS (C1  C2 )  He Abs Corr C2

91

(Eq. 2.3)

where FRC VS C1 C2

   

Functional residual capacity Volume in spirometer and circuit Concentration of He in spirometer Concentration of He at the end of the test after equilibration of lungs and spirometer He Abs Corr  Volume of theoretical He absorption by the blood during rebreathing (usually 0.1 liter); this is a controversial correction The volume in the spirometer and circuit (VS) includes the known volume of He and a dead space that consists of the tubing, the analyzer, and the space in the CO2 absorber. This dead space must be measured. In most automated systems, this is accomplished by initially adding a small amount of He to the closed circuit and taking an initial reading. A known volume of air is then added, and after a short time to allow equilibration, a second He concentration reading is taken. This additional step modifies equation 2.3. Because the initial volume in the spirometer and circuit (VS) can be calculated by taking another He concentration reading, the equation for FRC by He dilution then becomes: FRC 

C1 C2  C3   Vadded  He Abs Corr C3 C1  C2

(Eq. 2.4)

where FRC  Functional residual capacity C1  Concentration of He in spirometer and circuit at the start of the test C2  Concentration of He in spirometer and circuit after a known volume of air is added to estimate the spirometer and circuit dead space C3  Concentration of He in spirometer and circuit at the end of the test after equilibration of lungs and spirometer system Vadded  Volume of air added to spirometer and circuit between C1 and C2 He Abs Corr  Volume of theoretical He absorption by the blood during rebreathing (usually 0.1 liter); this is a controversial correction An example of calculating an FRC by closed-circuit He dilution (FRCHe) is as follows: Volume of air added Initial He concentration (C1) He concentration after air added (C2) He concentration after subject equilibration (C3) Spirometer temperature He absorption correction Switching error

c02.indd 91

      

1.20 liters 10.0% 8.6% 6.2% 23 C 0.1 liter 0

5/26/2011 6:29:31 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

92

CHAPTER 2

FRCHe 

Lung Volumes

C1 C2  C3   Vadded  He Abs Corr C3 C1  C2

10.0 8.6  6.2  1.200 @ ambient temperature and pressure, 6.2 10.0  8.6 satuurated with water vapor (ATPS)  0.10 FRCHe 

2.4  1.20  0.10 1.4 FRCHe  3.21  ATPS to BTPS factor of 1.084

FRCHe  1.61 

FRCHe  3.48 liters Testing Technique The testing technique for measuring FRCHe includes: (a) equipment preparation and calibration, (b) patient preparation and instruction, (c) testing, and (d) quality control. Equipment Preparation and Calibration Calibrate the spirometer used for He dilution systems as recommended by the manufacturer and test it for leaks every day it is used. The CO2 and H2O absorbers should be fresh. These absorbers usually contain color indicators. The CO2 absorber (soda lime) is usually white with color indicators that change to blue as they become saturated with CO2. The H2O absorber (calcium sulfate) is usually light blue with color indicators that change to pink as they become saturated with H2O. Have an adequate supply of He and O2 available. Patient Preparation and Instruction The patient should be seated comfortably and told how to use the mouthpiece and nose clip. Tell the patient that the test requires sitting quietly and breathing through the mouthpiece with the lips tightly sealed for several minutes. Remind the patient not to remove the mouthpiece until you say so. Because perforated eardrums can result in leaks to the system circuit and thus result in erroneous FRC values, ascertain if the patient has this condition. If so, supply earplugs. Testing With the patient wearing a nose clip and breathing room air while attached to the mouthpiece, observe tidal breathing for 30 to 60 seconds as the patient becomes accustomed to breathing on a mouthpiece. When a stable breathing pattern exists, turn the valve that connects the patient to the spirometer system and test gas at TV end-expiratory level (FRC). Often, the switch-in of the valve to connect the patient to the spirometer system is not done exactly at FRC. The patient may be a little above or a little below FRC. This is not a problem because a mathematical correction for this switching error is made in the calculations. Because it can be difficult to see the patient’s chest movement to turn him or her into the system exactly at FRC, try placing your hand on the patient’s back or shoulder to feel the chest movement and help determine the correct turn-in point. The calculation of FRCHe by the closed-circuit He dilution technique (equations 2.3 and 2.4) assumes that the volumes in the spirometer and lungs will not change between the start and end of the test. The CO2 produced by the patient will increase the spirometer volume.

c02.indd 92

5/26/2011 6:29:32 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of FRC

93

The O2 consumed by the patient will decrease the spirometer volume. Hence, CO2 must be absorbed, and O2 must be added. The CO2 produced by the patient is continually being absorbed by chemical absorption as the He mixture circulates in the spirometer, circuit, and patient. The O2 is added by one of two techniques. The first and most common technique is a continuous-addition technique, which keeps the spirometer system volume constant by continually adding O2 in small quantities that equal the volume loss caused by the patient’s O2 consumption. Typically, an adult individual on this type of system has an O2 consumption of 300 to 600 mL/min. The second O2 addition technique, found on some older systems, is the bulk-addition technique. Systems that use this technique put a bolus of O2 (e.g., 1000 mL) into the spirometer before the patient is turned in. When the patient is turned in, all of the added O2 must be consumed before the test can conclude. Otherwise, the volume of the spirometer and circuit will have changed, causing a greater He concentration change than would normally be seen. The patient should continue to breathe on the mouthpiece until equilibration is reached. The He concentration is monitored every 15 seconds. Equilibration is usually defined by a plateau in the He concentration (i.e., the change in He concentration is less than 0.02% over 30 seconds).2 If the He concentration continues to fall steadily with no trend toward equilibration, suspect a leak. Leaks typically occur around the patient’s mouth or from a loose nose clip. If equilibration has not been reached 10 minutes after the beginning of the test, end the test. For patients who may become hypoxic, use pulse oximetry monitoring. At the end of the test (i.e., equilibration), perform an SVC maneuver before removing the mouthpiece (linked to FRC determination); some patients, however, may not be able to perform this additional test after being on the mouthpiece for several minutes. If the patient cannot perform the linked SVC maneuver before being disconnected, have the patient perform it as soon as possible after FRC determination and disconnection. When the patient comes off the mouthpiece, saliva and secretions spill out, so have a good supply of tissues nearby. Perform at least one technically acceptable FRCHe determination. If more than one determination is performed, allow at least 5 minutes between them.2,21 For practical purposes, differences of less than 10% between two FRCHe determinations are acceptable.2,22 Report the mean FRCHe in liters at BTPS rounded to two decimals (e.g., 3.13 liters). Quality Control The accuracy of the He dilution device can be determined by using a 3-liter calibration syringe. The He dilution system should be prepared to test a patient. Connect the 3-liter syringe and inject a known amount of air (e.g., 3 liters) into the spirometer. Close the valve to the syringe, if possible, to avoid mixing the dead space of the syringe after injecting the volume of air. After approximately 30 seconds, when the meter reading is stable, end the test. The calculated volume (at ATPS) should equal the injected volume within 3%. If this is not the case, evaluate the system. Perform this quality control check at least weekly.12 In addition, the measurement of values in biological controls is useful and assures the testing values are stable. At least monthly, test two or three reference individuals (i.e., nonsmoking and healthy) on the system. Values obtained for FRCHe that differ by more than 10% from previously established means for a given individual should alert you to investigate the system.

c02.indd 93

5/26/2011 6:29:32 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

94

CHAPTER 2

Lung Volumes

Measuring FRC by Multiple-Breath Open-Circuit N2 Washout The third method of determining FRC is the multiple-breath open-circuit N2 washout (FRC N 2). This method involves removing or washing out the N2 gas in a patient’s lungs while the patient breathes 100% O2 for several minutes. It is easy for the patient to perform and requires minimal learning and effort. Like the He dilution method, the N2 washout method has the drawback of measuring only the lung spaces that communicate with the mouth. Thus it also underestimates FRC in individuals with airflow obstruction. The discussion of this method includes: (a) physiology and instrumentation and (b) testing technique. Physiology and Instrumentation The early modern technique23 for the open-circuit method used the apparatus shown in Figure 2.15. This technique is cumbersome compared to today’s techniques. It involved acquiring three gas samples (an alveolar sample prior to starting the test, a second alveolar sample after the test, and a sample of the exhaled gas during the 100% O2 breathing period). The first sample was taken with the patient attached to the system mouthpiece and breathing room air. The patient was instructed to exhale maximally, at which time the alveolar gas sample was taken, which represented the average lung N2 concentration on room air. Following this sample, the patient breathed room air for 2 more minutes to restore quiet breathing. Then a valve was turned at FRC so the patient could breathe 100% O2 from a reservoir bag for 7 minutes. The exhaled gas was collected in a 100-liter gasometer (Tissot). After 7 minutes, the patient was instructed to exhale fully, and the second alveolar sample was taken. The patient was then disconnected, the breathing circuit was flushed, and the third gas sample was taken from the gasometer. The three N2 samples were analyzed using the volumetric methods (e.g., Van Slyke manometer), which depended on chemical absorption. This analyzation process, which was common in the early 1960s, took 10 to 20 minutes for each sample and was extremely technique dependent. Today, the large gasometer has been replaced by flow meters, and the volumetric gas analyzers have been replaced by electronic N2 analyzers that can analyze the N2 concentration of each breath. Additionally, the second alveolar sample at the end of the O2 breathing period has been eliminated from the technique.24 The concept of the N2 washout is based on the fact that at the start of the test the unknown FRC contains 80% N2 and an unknown concentration of O2 (probably between 16% and 21%)

Figure. 2.15 The early modern open-circuit N2 washout circuit for determining FRC. The patient breathes through a two-way valve to which a vacuum bottle (A) is connected for collecting the alveolar sample. The inspiratory side is connected to a 100% oxygen reservoir, and the expiratory side is connected to a large water-sealed spirometer (T). This spirometer, which is also known as a Tissot (pronounced TEE-so), is usually 100 to 200 liters, which is large enough to collect the expired air during 7 minutes of quiet breathing.

c02.indd 94

5/26/2011 6:29:32 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of FRC

95

Two-way valve

A

T

100% oxy

gen

c02.indd 95

5/26/2011 6:29:32 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

96

CHAPTER 2

Lung Volumes

and CO2 (probably between 0.4% and 5.0%). By measuring the volume of N2 in the FRC and applying a concentration dilution formula, the FRC volume can be determined. Thus, C1V1  C2V2 where C1  N2 concentration at start of test V1  FRC volume C2  N2 concentration in exhaled volume V2  Total exhaled volume during O2 breathing period As stated earlier, the N2 washout measures the lung volume that can be ventilated by the mouth. Thus, it will underestimate FRC in those individuals with large amounts of trapped air or areas of lung that are poorly ventilated. Early data23,25 showed that increased breathing periods (i.e., 11 to 15 minutes) obtained larger FRCs in individuals with obstructive lung disease. However, longer durations are more uncomfortable for the patient, and in some patients (e.g., those with COPD) who breathe 100% O2 for more than a few minutes, it may result in CO2 retention.26 The measurement of N2 can be performed using N2 analyzers (most common), mass spectrometers, or indirectly by subtracting measurements of O2 and CO2. The N2 analyzers used with modern systems operate on a photoelectric principle. Gas is pulled through a needle valve into an ionization chamber by a vacuum pump. The molecules are ionized and emit light. This light is then filtered and collected by a photo resistor, which converts the light into an electrical signal. The intensity of the light is directly related to the concentration of N2 in the sample. The use of computers has allowed for the signal from the N2 analyzer to be combined with the volume signal from a spirometer to provide instantaneous or breath-by-breath measurements of N2 and volume. This technology allows for faster detection of leaks, which is an advantage over the He dilution technique. The basic equation used with the early technique was: FRCN2 

(Tissot volume) (FEN 2  FIN 2 )  DS FAN 2 initial  FAN 2 final

An example of calculating an FRC by N2 washout using the previous equation is as follows. An open-circuit N2 washout was performed with the patient inspiring 100% O2, and the exhaled air was collected in a 120-liter Tissot gasometer. The following information was obtained: Barometric pressure (PB) Tissot temperature (T) Tissot volume after 7 minutes of breathing FEN2 (fractional concentration of expired N2 in Tissot) FIN2 (fractional concentration of inspired N2) FAN2 initial (alveolar N2 concentration at start of test) FAN2 final (alveolar N2 concentration at end of test) DS (valve dead space, liters)

c02.indd 96

       

631 mmHg 24 C 56.3 liters ATPS 0.0368 0.001 0.80 0.015 0.090

5/26/2011 6:29:32 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of FRC

FRCN2 

97

(Tissot volume) (FEN 2  FIN 2 )  DS FAN 2 initial  FAN 2 final

VE Tissot (i.e., Tissot volume @ ATPS including system dead space)  56.3 liters VE Tissot PB  PH 2 O at 24°C 310  VE Tissot @ BTPS   @ ATPS 2773  T PB  PH 2 O at 37°C VE Tissot @ BTPS  56.3 

310 631  22  273 + 24 631  47

VE Tissot @ BTPS  61.30 liters (61.30)(0.0368  0.001) FRCN2  0.80  0.015 (61.30)(0.0358)  2.80 liters @ BTPS FRCN2  0.785 Today, the computerization and simplification of this technique has greatly shortened the time needed to calculate FRCN2. Testing Technique The testing technique for measuring FRCN2 includes: (a) equipment preparation and calibration, (b) patient preparation and instruction, (c) testing, and (d) quality control. Equipment Preparation and Calibration Calibrate the flow-meter device and N2 analyzer every day of use, following the manufacturers instructions. There should be an ample supply of O2. Patient Preparation and Instruction The patient should be seated comfortably and given instructions on how to use the mouthpiece and nose clip. Tell the patient that the test requires sitting quietly and breathing on the mouthpiece with the lips tightly sealed for several minutes. Remind the patient to keep the mouthpiece in place with the lips sealed until you say to remove it. Because perforated eardrums can result in leaks to the system circuit and thus result in erroneous FRC values, ascertain if the patient has this condition. If so, supply earplugs. Testing The patient breathes on the mouthpiece for 30 to 60 seconds to become comfortable on the mouthpiece and to assure a stable FRC level. When a stable FRC level has been established, the patient is turned in to the system and starts breathing 100% O2. The N2 concentration and the patient are monitored during the washout period to assure that no leaks occur. A large increase in N2 concentration indicates a leak. If this occurs, the test should be stopped and repeated after the appropriate waiting period (i.e., 15 minutes). The test is usually stopped when the expired N2 concentration falls below 1.5% for at least three successive breaths.2

c02.indd 97

5/26/2011 6:29:32 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

98

CHAPTER 2

Lung Volumes

Monitor the N2 concentration throughout the test. When no leaks are present, the N2 concentration appears as shown in Figure 2.16A. If a leak occurs, N2 from room air enters the system, and the N2 concentration abruptly rises (Figure 2.16B).

Figure 2.16 A. A typical display of an N2 washout test using an N2 analyzer. B. When leaks occur (e.g., the patient does not keep the lips sealed tightly), room air enters the circuit, and the N2 concentration spikes to near ambient levels (i.e., near 80%). 90

A

80 70 60 N2

50 40 30 20 10 0 Time 90

B

80 70

leak

60 N2

50 40 30 20 10 0 Time

c02.indd 98

5/26/2011 6:29:33 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Measurement of Lung Volumes by Other Methods

99

At least one technically satisfactory measurement should be obtained. If additional washouts are performed, allow at least 15 minutes between maneuvers. Repeat the N2 washout procedure with a 15-minute interval between trials until you obtain two FRC measurements that agree within 10%.2 The mean FRCN2 measurements that agree within 10% should be reported in liters at BTPS, rounded to two decimals (e.g., 3.78 liters). Quality Control The main cause of erroneous results with the open-circuit N2 washout test is leaks. These can be easily detected during the test by observing the N2 signal. As with the He dilution and body box techniques, the testing of biological controls is recommended at least monthly.

Measurement of Lung Volumes by Other Methods There are other less commonly used methods for measuring lung volumes, and these will only be briefly discussed. The single-breath N2 method can be used to estimate RV and TLC.21 The test involves diluting the ambient N2 in the lungs by inhaling a VC (from RV to TLC) of 100% oxygen. An analysis of the single-breath N2 curve produced in this test yields several other parameters of pulmonary function, including closing volume and slope of phase III. However, this method has never been standardized and is not widely used. The single-breath He method estimates TLC and is performed during the single-breath carbon-monoxide uptake (DL,CO) (discussed in Chapter 3). This DL,CO test employs an inert tracer gas (e.g., He), which is diluted into the lung volume, and an estimate of TLC can be calculated. This estimate of TLC is referred to as alveolar volume (VA) and is lower than the true TLC in individuals with airflow obstruction. The TLC can also be estimated from the chest roentgenogram (radiograph) using the posteroanterior (PA) and lateral (LAT) views when the patient is at full lung inflation. The two basic methods that use the chest radiograph to measure lung volumes are (a) the planimeter method27,28 and (b) the ellipse method.29 A planimeter is a device with two arms that pivot around a fixed joint, allowing one to trace over the lines of any two-dimensional shape (Figure 2.17). By outlining all the lung fields on the PA and LAT films, including the heart but excluding the sternum, the radiographic chest volume can be calculated. Digital technology has been developed to replace the manual method. The ellipse method assumes that the lungs can be divided into a large number of elliptical cross sections (Figure 2.18). The area of the elliptical sections is determined from the PA and LAT view chest films and then converted to volume. The area of the heart, the domes of the diaphragm, and the pulmonary blood and tissue are subtracted out. Both methods compare well with each other as well as with TLC measured by plethysmographic and dilution–washout techniques.30-33 The drawbacks of using the chest film are: (a) only TLC can be measured and (b) TLC can be underestimated if the patient does not maximally inspire when the film is exposed. Additionally, the technique depends on good-quality film to define the lung boundaries.

c02.indd 99

5/26/2011 6:29:33 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

100

CHAPTER 2

Lung Volumes

Figure 2.17 Measurement of TLC from chest roentgenograms using the planimetry technique. The twoarmed planimeter is placed on the PA and LAT view of the chest at full inspiration. With one end fixed, the other end is used to trace the periphery of each view. A mechanical counter measures the distance traveled, which can be converted to volume.

PA view

910

LAT view

23

23 910

Planimeter

However, the advantages of being unaffected by poorly ventilated areas and the ability to compare films retrospectively when pulmonary function tests are not available or comparable make this technique valuable. The use of computers has shortened the time required to measure and calculate the data. CT scans can provide estimates of lung volumes and can estimate the volume of the lung with increased or decreased density.34 Magnetic resonance imaging (MRI) offers the advantage of scanning specific regions of the lung. However, MRIs are costly, so their use for measuring lung volumes is limited.

Reference Values As in forced spirometry, it is common practice to report reference or predicted values for static lung volumes. Unfortunately, no one set of lung volume reference values was recommended by the most recent guidelines.2 One survey35 found that the most commonly used study for lung volume reference values was that of Goldman and Becklake.36 This study used hydrogen (H) dilution (similar to He dilution) to measure FRC on 44 male and 50 female subjects near Johannesburg, South Africa (altitude 5,700 feet).

c02.indd 100

5/26/2011 6:29:33 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Infection Control

101

Figure 2.18 Measurement of TLC from chest roentgenograms using the ellipse method. The PA and LAT views of the chest at full inspiration are divided into five elliptical cross sections or segments. The area of each section is measured (arrows indicate the points between which to measure) and then converted to volume.

PA view

LAT view

I

I

II

II

III

III

IV IV

V

V

The techniques used in the many reference studies vary and include He dilution, H dilution, N2 washout, single-breath N2, and body plethysmography.37-41 This fact complicates the task of selecting reference equations for one’s laboratory. When selecting reference values, consider the following criteria: (a) use of similar equipment and methods and (b) similar populations (i.e., age, body size, gender, and race). When a tentative selection has been made, a laboratory should consider comparing the measurements from 10 to 20 healthy individuals selected from a representative sample of the laboratory’s population with the tentative reference values. If the differences between the values of the 10 to 20 healthy individuals and reference values are small (e.g., within ± 10%), then the laboratory can be reasonably confident that the chosen reference values are appropriate. If the differences are greater, consider using other reference values.

Infection Control In determining FRC and the lung volume compartments, the patient continually rebreathes on the breathing circuit. Thus, the possibility of cross contamination exists. The laboratory should have procedures in place to help ensure patient and technologist safety. The Centers for Disease Control and Prevention has published guidelines for preventing transmission of infectious agents in healthcare settings.42 In addition, Kendrick and coworkers published an excellent review and practical approach to infection control in the pulmonary function laboratory.43

c02.indd 101

5/26/2011 6:29:33 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

102

CHAPTER 2

Lung Volumes

If the patient has known Mycobacterium tuberculosis and is considered infectious, several additional precautions are recommended. Test the patient at the end of the day to ensure no other patient uses the instrument that day. The room in which the test is performed should at least meet current Centers for Disease Control and Prevention guidelines for air changes and ventilation. The airflow should be biased into the contaminated area from adjacent hallways and rooms (i.e., negative pressure). The use of ultraviolet light is controversial, but some centers use it. The technologist performing the testing should wear a personal respirator that meets current Occupational Safety and Health Administration (OSHA) recommendations. After testing is complete, disassemble the tubing and removable parts from the testing system and clean them with a high-level disinfectant. Reassemble the system the next day after all the parts have thoroughly dried.

Case Presentations Case 2.1 A 68-year-old Caucasian woman with a complaint of worsening dyspnea was seen in the pulmonologist’s office. She admitted to having been a heavy smoker (90 pack years, where one pack year is defined as one package of cigarettes per day for 1 year) but had recently quit. This office was equipped to measure spirometry and lung volumes, and the results of her tests are shown in Table 2.2 and Figure 2.19. Her chest film had an emphysematous appearance with increased diameters. After seeing the results of the office pulmonary function test, the doctor ordered another pulmonary function test at the area hospital, and the results are shown in Table 2.3 and Figure 2.20.

Table 2.2 Office Pulmonary Function Data Before and After a Bronchodilator Predicted

Before*

4.82

4.63 (96)

FRCHe (L)

2.80

2.58 (92)

RV (L)

1.91

2.08 (109)

SVC (L)

3.59

2.55 (71)

FVC (L)

3.63

2.18 (60)

2.38

FEV1 (L)

2.57

1.08 (42)

1.18

TLC (L) †

FEV1/FVC (%) FEF25–75% (L/sec)

71 2.50

50 0.65 (26)

After

50 0.67

*Values in parentheses are percent predicted †

Measured by He dilution

c02.indd 102

5/26/2011 6:29:33 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Case Presentations

103

Figure 2.19 Volume–time spirogram from patient in Case 2.1 performed in the doctor’s office before and after administration of a bronchodilator. Before After

5

4

3 Volume (L) 2

1

0 2

4

6

8

10

12

14

Time (sec)

Table 2.3 Hospital Pulmonary Function Laboratory Data Before and After a Bronchodilator Predicted TLC (L)

Before*

After

4.82

5.76 (120)

FRCpleth (L)

2.80

3.97 (142)

RV (L)

1.91

3.13 (164)

SVC (L)

3.59

2.63 (73)

2.68

FVC (L)

3.63

2.28 (63)

2.43

FEV1 (L)

2.57

1.11 (43)

1.29



FEV1/FVC (%) FEF25–75%(L/sec)

71

49

53

0.59 (24)

0.83

Raw (cm H2O/L/sec)

2.93

1.84

sGaw (L/cm H2O/L/sec)

0.07

0.13

DL,CO (mL/min/mmHg) DL,CO/VA

2.50

23.4 4.85

17.30 (74) 3.64 (75)

*Values in parentheses are percent of predicted †

Measured by body plethysmograph

c02.indd 103

5/26/2011 6:29:34 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

104

CHAPTER 2

Lung Volumes

Figure 2.20 Flow–volume spirogram from patient in Case 2.1 performed in the hospital pulmonary function laboratory showing before and after administration of a bronchodilator, as well as the reference (i.e., predicted) curve. 10

8

Predicted After Before

6 Flow (L/sec) 4

2

0 1

2

3

4

5

6

Volume (L)

Questions 1. 2. 3.

How would you interpret the pulmonary function tests done in the doctor’s office? How would you interpret the pulmonary function tests done in the hospital? How do you explain the difference in data?

Answers and Discussion Data from the pulmonary function test done in the office and shown in Table 2.2 reveal normal lung volumes (TLC, FRCHe, and RV) and a reduced SVC. These data also reveal severe airflow limitation with minimal response to a bronchodilator. A concern is why the lung volumes do not reveal hyperinflation that would be consistent with this degree of obstruction and the chest film interpretation. Therefore, the final interpretation would most likely state a mixed obstructive and restrictive disorder. The pulmonary function test data from the hospital (Table 2.3), which used the body plethysmograph, show a different picture. There is still severe airflow limitation and the diffusing capacity is reduced, but the lung volumes are all markedly increased. There is good response to a bronchodilator. The final interpretation from these data is severe airflow limitation, but there is certainly not a mixed disorder.

c02.indd 104

5/26/2011 6:29:34 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Case Presentations

105

The major difference between the two interpretations depends on the measurement of FRC. In the doctor’s office, the FRCHe was 2.55 liters. In the hospital laboratory, the FRCpleth was 3.97 liters, which is significantly higher. Why are the two FRC determinations so different? Measurement of FRCHe uses a gas dilution technique. It requires the patient to breathe into a system that contains a known volume and concentration of He. During the breathing period, the He diffuses and equilibrates into the gas spaces of the lung that communicate with the mouth. Noncommunicating spaces are not measured. Thus, the FRCHe can be underestimated. The body plethysmograph (body box) is another method to determine FRC. Although it is more expensive and requires more technical expertise, it is quick and accurate. It measures the TGV in the thoracic cage when a shutter is closed. This volume includes non-communicating areas and therefore is more accurate than the gas dilution technique. The difference between the two sets of lung volumes in this patient can be accounted for by the presence of trapped gas spaces or poorly communicating airways. The correct interpretation is that the patient is hyperinflated and does not have a mixed disorder.

Case 2.2 A 58-year-old African American male office worker was tested in the pulmonary function laboratory. Spirometry at a recent physical examination in the doctor’s office detected some abnormal results. He denied shortness of breath or cough but claimed to be a 40 pack year smoker. His laboratory results are shown in Table 2.4 and Figure 2.21.

Table 2.4 Pulmonary Function Values Before and After a Bronchodilator in the Pulmonary Function Laboratory

FVC (L) FEV1 (L) FEV1/FVC (%)

Predicted

Before*

After

5.04

3.78 (75)

3.93

3.51 70

2.00 (57) 53

FEF25–75% (L/Sec)

3.97

2.34 (59)

FRCpleth (L)

4.01

2.89 (72)

RV (L)

1.88

1.94 (103)

TLC (L)

6.88

5.85 (85)

SVC (L)

5.01

3.91 (78)

2.12 54 2.79

Raw (cm H2O/L/sec)

2.13

1.92

sGaw (L/cm H2O/L/sec)

0.15

0.19

*Values is parentheses are percent predicted

c02.indd 105

5/26/2011 6:29:34 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

106

CHAPTER 2

Lung Volumes

Figure 2.21 Flow–volume spirogram from patient in Case 2.2 showing before and after bronchodilator curves as well as the reference (i.e., predicted) curve. 10

8

Predicted After Before

6 Flow (L/sec) 4

2

0 1

2

3

4

5

6

Volume (L)

Questions 1. 2.

What is the interpretation and evaluation of these pulmonary function tests? What issues should be raised with the reference values?

Answers and Discussion The data from Table 2.4 and Figure 2.21 reveal airflow limitation without meaningful response to a bronchodilator. Additionally, the lung volumes are reduced, suggesting a mixed obstructive and restrictive disorder. The predicted or reference values used by the laboratory do not state whether they are race specific. If they are based on a Caucasian population, then the apparent restrictive process could be an artifact. Healthy, predicted, or normal values for pulmonary function tests are frequently based on entirely Caucasian populations. The predicted values for the lung volumes (TLC, RV, FRC) shown in Table 2.4 come from a study by Goldman and Becklake,36 and the other predicted values come from a study by Morris and coworkers.44 Neither of these particular studies state the race or ethnic background of the subjects who participated. Several studies have documented that people of African descent have lower lung volumes than do Caucasians.45-49 This is attributed to the fact that people of African descent have longer

c02.indd 106

5/26/2011 6:29:34 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Case Presentations

107

legs and shorter trunks. Thus, standing height, which is a major factor in predicting pulmonary function values, is biased. When the predicted values are recalculated (shown in Table 2.5) based on the studies of Rossiter and Weil47 and Stinson and colleagues,48 which used black populations, the interpretation changes to airflow limitation. In practice, some pulmonary function laboratories use scaling factors when testing African Americans. The process adjusts the Caucasian population’s predicted values down approximately 12%. This approach has a major pitfall in that the 12% correction is an average. The difference between Caucasian and African American values varies with each parameter. For TLCs, for example, African Americans have approximately 11% lower predicted values.47 A pulmonary function laboratory must carefully consider selecting predicted values. The use of race-specific predicted equations raises the issue of whether the available studies are adequate, because they do not provide criteria for determining race. Many studies assume that racial identity is evident through color distinctions. However, this may not always be true in practice. The alternative to race-specific predicted values is to take racial issues into consideration at interpretation. For example, in this case, the interpretation of the data in Table 2.4 might have read “airflow limitation without meaningful response to a bronchodilator and normal lung volumes when adjusted for race.” The important consideration is that there is a weakness in using healthy reference equations based on Caucasian populations for non-Caucasian patients. If race-specific reference equations are used, note so on the report.

Table 2.5 Pulmonary Function Values Before and After Bronchodilator Predicted

Before*

After

FVC (L)

3.98

3.78 (95)

3.93

FEV1 (L)

2.82

2.00 (71)

2.12

FEV1/FVC (%)

71

53

FEF25–75% (L/Sec)

3.60

2.34 (65)

FRCpleth (L)

3.28

2.89 (88)

RV (L)

2.04

1.94 (95)

TLC (L)

6.03

5.85 (97)

SVC (L)

3.99

54 2.79

3.91 (98)

Raw (cm H2O/L/sec)

2.13

1.92

sGaw (L/cm H2O/L/sec)

0.15

0.19

*Values in parentheses are percent predicted

c02.indd 107

5/26/2011 6:29:34 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

108

CHAPTER 2

Lung Volumes

Self-Assessment Questions 1.

2.

3.

4.

5.

6.

c02.indd 108

All the following are techniques for determining FRC except: a. Closed-circuit He dilution b. Open-circuit N2 washout c. Single-breath N2 washout d. Body plethysmography Which of the following is the most accurate concerning the ERV? a. The maximum amount of air that can be exhaled from the VC b. The maximum amount of air that can be exhaled from the TV end-expiratory level c. The maximum amount of air that can be inhaled from the TV end-expiratory level d. The maximum amount of air that can be exhaled from RV A patient has the following lung volumes: Observed

Predicted

% Predicted

SVC (L)

3.50

4.30

81

FRCpleth (L)

3.80

3.00

127

RV (L)

3.00

2.00

150

TLC (L)

6.50

6.30

103

The interpretation would most likely state: a. Normal lung volumes b. Hyperinflation c. Restrictive pattern d. Mixed obstructive and restrictive pattern In patients with obstructive lung disease, the gas dilution methods for FRC determination: a. Overestimate FRC b. Equal body plethysmograph FRC c. Underestimate FRC d. Equal radiographic FRC In the body plethysmograph, alveolar pressure changes caused by the compression and decompression of the lungs are estimated by: a. Measuring mouth pressure b. Measuring body box pressure c. Measuring transpulmonary pressure d. Measuring transdiaphragmatic pressure The calculation of FRC using the body plethysmograph is based on: a. Murphy’s law b. Boyle’s law c. Charles’s law d. Poiseuille’s law

5/26/2011 6:29:34 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

Self-Assessment Questions

7.

8.

9.

10.

11.

12.

13.

14.

c02.indd 109

109

Which of the following methods for determining TLC best agrees with the results of the radiographic technique in patients with obstructive lung disease? a. Body box FRC  IC b. Closed-circuit He dilution FRC  IC c. Open-circuit N2 washout FRC  IC d. Open-circuit He dilution FRC  IC Which of the following is most accurate about the FRC? a. It is the volume of air remaining in the lungs at TV end-expiratory level b. It consists of residual volume and expiratory reserve volume c. It can be determined by gas dilution and body plethysmography d. All of the above e. a and c In restrictive lung disease, which lung volume compartment is not decreased? a. VC b. FRC c. TLC d. IC e. None of the above In obstructive lung disease, the SVC will probably be: a. Larger than the TLC b. Smaller than the RV c. Larger than the FVC d. Smaller than the IC Lung volume values for TLC, FRC, and RV should be reported at: a. BTPS b. ATPS c. STPD (standard conditions) d. ATPD (ambient temperature and pressure, dry) Which of the following statements best explains why it is important to wait at least 5 minutes between He dilution FRC determinations? a. The subject must recover from breathing on the mouthpiece b. The subject’s CO2 production must return to resting levels c. The He must be cleared from the subject’s lungs d. To ensure that O2 saturation has returned to baseline Which of the following should be done to ensure that the volume in the spirometer during a closed-circuit FRC determination remains unchanged? a. Ensure adequate He absorption b. Ensure adequate CO2 absorption c. Ensure adequate O2 absorption d. Ensure subject starts test at FRC When the N2 concentration abruptly rises near the end of an open-circuit N2 washout, it usually means: a. The computer has an error b. The N2 analyzer is not calibrated

5/26/2011 6:29:34 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

110

CHAPTER 2

Lung Volumes

c. The system has developed a leak d. The O2 concentration is varying 15. Which of the following are methods to estimate TLC? a. Single-breath N2 b. Single-breath He c. Measurement of chest radiographs d. a, b, and c e. a and c

References 1. Miller MR, Hankinson J, Brusasco V, et al. Standardisation of spirometry. Eur Respir J. 2005;26:319–338. 2. Wanger J, Clausen JL, Coates A, et al. Standardisation of the measurement of lung volumes. Eur Respir J. 2005;26:511–522. 3. Tierney DF, Nadel JA. Concurrent measurements of functional residual capacity by three methods. J Appl Physiol. 1962;17:871–873. 4. Reichel G. Differences between intrathoracic gas measured by the body plethysmograph and functional residual capacity determined by gas dilution methods. Prog Respir Res. 1968;4:188–193. 5. Cobeel LJ. Comparison between measurement of functional residual capacity and thoracic gas volume in chronic obstructive pulmonary disease. Prog Respir Res. 1969;4:194–204. 6. DuBois AB, Botelho SV, Bedell GN, et al. A rapid plethysmographic method for measuring thoracic gas volume: a comparison with a nitrogen washout method for measuring FRC in normal subjects. J Clin Invest. 1956;35:322–326. 7. Rodenstein DO, Stanescu DC. Reassessment of lung volume measurement by helium dilution and body plethysmography in chronic airflow obstruction. Am Rev Respir Dis. 1982;126:1040–1044. 8. Rodenstein DO, Stanescu DC, Francis C. Demonstration of failure of body plethysmography in airway obstruction. J Appl Physiol. 1982;52:949–954. 9. Shore SA, Huk O, Mannix S, Martin JG. Effect of panting frequency on the plethysmographic determination of thoracic gas volume in chronic obstructive pulmonary disease. Am Rev Respir Dis. 1983;128:54–59. 10. Shore S, Milic-Emili J, Martin JG. Reassessment of body plethysmographic technique for the measurement of thoracic gas volume in asthmatics. Am Rev Respir Dis. 1982;126:515–520. 11. O’Donnell CR, Bankier AA, Stiebellehner L, et al. Comparison of plethysmograph and helium dilution lung volumes: which is best for COPD? Chest. 2010;137:1108–1115. 12. Quanjer PH, Tammeling GJ, Cotes OF, et al. Lung volumes and forced ventilatory flows. Report working party. Standardization of lung function tests, European Community for Steel and Coal. Eur Resp J. 1993;6(suppl 16):5–40. 13. Comroe J. Retrospectroscope. Man-cans. Am Rev Respir Dis. 1977;116:945–950.

c02.indd 110

5/26/2011 6:29:35 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

References

111

14. Coates AL, Desmond KJ, Demizio D, et al. Sources of error in flow–volume curves. Effect of expired volume measured at the mouth vs that measured in a body plethysmograph. Chest. 1988;94:976–982. 15. Sauder LR. Computer analysis versus technician analysis of body plethysmographic analog recordings of airway resistance and thoracic gas volume. Respir Care. 1982;27:62–69. 16. Clausen JL, Zarins LP. Pulmonary function testing guidelines and controversies. New York, NY: Academic Press; 1982. 17. Coates AL, Peslin R, Rodenstein D, Stocks J. Measurement of lung volumes by plethysmography. Eur Respir J. 1997;10:1415–1427. 18. Meneely GR, Ball COT, Kory RC, et al. A simplified closed circuit helium dilution method for the determination of the residual volume of the lungs. Am J Med. 1960;28:824–831. 19. Hathirat S, Renzetti AD, Mitchell M. Measurement of the total lung capacity by helium dilution in a constant volume system. Am Rev Respir Dis. 1970;102:760–770. 20. Brown R, Enright P, Leith D. Multiple-breath helium dilution measurements of lung volumes in adults. Eur Respir J. 1998;11:246–255. 21. Schaanning CG, Gulsvik A. Accuracy and precision of helium dilution technique and body plethysmography in measuring lung volumes. Scand J Clin Lab Invest. 1973;32:271–277. 22. Darling RC, Cournand A, Richards DW Jr. Studies on the intrapulmonary mixture of gases. III. An open circuit method for measuring residual air. J Clin Invest. 1940;19:609–618. 23. Emmanuel G, Briscoe WA, Cournand A. A method for the determination of the volume of air in the lungs: measurements in chronic pulmonary emphysema. J Clin Invest. 1961;20:329–337. 24. Newth CJ, Enright P, Johnson RL. Multiple breath nitrogen washout techniques: including measurements with patients on ventilators. Eur Respir J. 1997;10:2174–2185. 25. Buist AS, Ross BB. Predicted values for closing volumes using a modified single-breath nitrogen test. Am Rev Respir Dis. 1973;107:744–752. 26. Hansen DW, Marshall BE, Frasch HF, Marshall C. Causes of hypercarbia with oxygen therapy in patients with chronic obstructive pulmonary disease. Crit Care Med. 1996;24:23–28. 27. Hurtado A, Fray WN. Studies of total pulmonary capacity and its subdivisions. II. Correlation with physical and radiologic measurements. J Clin Invest. 1933;12:807–823. 28. Pratt PC, Klugh GA. A method for the determination of total lung capacity from posteroanterior and lateral chest roentgenograms. Am Rev Respir Dis. 1967;96:548–552. 29. Barnhard HJ, Pierce JA, Joyce JW, Bates JH. Roentgenographic determination of total lung capacity. Am J Med. 1960;28:51–60. 30. Harris TR, Pratt PC, Kilburn KH. Total lung capacity measured by roentgenograms. Am J Med. 1971;50:756–763. 31. Ries AL, Clausen JL, Friedman PJ. Measurement of lung volumes from supine portable chest radiographs. J Appl Physiol. 1979;47:1332–1335. 32. Pierce RJ, Brown DJ, Holmes M, et al. Estimation of lung volumes from chest radiographs using shape formation. Thorax. 1979;34:726–734. 33. Miller RD, Offord KP. Roentgenologic determination of total lung capacity. Mayo Clin Proc. 1980;55:694–699.

c02.indd 111

5/26/2011 6:29:35 PM

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION

112

CHAPTER 2

Lung Volumes

34. Clausen JL. Measurement of absolute lung volumes by imaging techniques. Eur Respir J. 1997;10:2427–2431. 35. Ghio AJ, Crapo RO, Elliott CG. Reference equations used to predict pulmonary function. Chest. 1990;97:400–403. 36. Goldman HI, Becklake MR. Respiratory function tests: normal values at median altitudes and the prediction of normal results. Am Rev Tubercul. 1959;79:457–467. 37. Grimby G, Soderholm B. Spirometric studies in normal subjects. III. Static lung volumes and maximum voluntary ventilation in adults with a note on physical fitness. Acta Med Scand. 1963;173:199–206. 38. Boren HG, Kory RC, Syner JC. The Veterans Administrative–Army cooperative study of pulmonary function. II. The lung volume and its subdivisions in normal men. Am J Med. 1966;41:96–114. 39. Crapo RO, Morris AH, Clayton PD, Nixon CR. Lung volumes in healthy nonsmoking adults. Bull Eur Physiopathol Respir. 1982;18:419–425. 40. Viljanen A, ed. Reference values for body plethysmography in non-smoking, healthy adults. Scand J Clin Lab Invest. 1982;42(suppl):35–50. 41. Stocks J, Quanjer PH. Reference values for residual volume, functional residual capacity, and total lung capacity. ATS Workshop on Lung Volume Measurements. Official Statement of the European Respiratory Society. Eur Respir J. 1995;8:492–506. 42. Siegel JD, Rhinehart E, Jackson M, Chiarello L; and Healthcare Infection Control Practices Advisory Committee. 2007 guideline for isolation precautions: preventing transmission of infectious agents in healthcare settings. http://www.cdc.gov/hicpac/2007IP/2007isolation Precautions.html. Accessed January 2011. 43. Kendrick AH, Johns DP, Leeming JP. Infection control of lung function equipment: a practical approach. Respir Med. 2003;97:1163–1179. 44. Morris JF, Koski A, Johnson LC. Spirometric standards for healthy nonsmoking adults. Am Rev Respir Dis. 1971;103:57–67. 45. Hsu KHK, Jenkins DE, Hsi BP, et al. Ventilatory functions of normal children and young adults—Mexican-American, white, and black. Spirometry. Pediatrics. 1979;95:14–23. 46. Schoenberg B, Beck GJ, Bouhuys A. Growth and decay of pulmonary function in healthy blacks and whites. Respir Physiol. 1978;33:367–393. 47. Rossiter CE, Weil H. Ethnic differences in lung function: evidence for proportional differences. Int J Epidemiol. 1974;3:55–61. 48. Stinson JM, McPherson GL, Hicks K, et al. Spirometric standards for healthy black adults. J Natl Med Assoc. 1981;73:729–733. 49. Boggs PB, Stephens AL, Walker RF, et al. Racially specific reference standards for commonly performed spirometric measurements for black and white children, ages 9–18 years. Ann Allergy. 1981;47:273–277.

c02.indd 112

5/26/2011 6:29:35 PM

Loading...

Lung Volumes - Jones & Bartlett Learning

© Jones & Bartlett Learning, LLC. NOT FOR SALE OR DISTRIBUTION Two CHAPTER 2 Lung Volumes  Introduction 69  Lung Subdivisions: Volumes and Cap...

2MB Sizes 6 Downloads 12 Views

Recommend Documents

Jones & Bartlett Learning Blog
Mary Elizabeth O'Brien's Spirituality in Nursing: Standing on Holy Ground, Sixth Edition Receives a 5-Star Review ... In

Descriptive Statistics - Jones & Bartlett Learning
Explain the purpose of descriptive statistics. • Compute measures of central tendency. • Compute measures of variabi

Chapter 4 - Jones & Bartlett Learning
every gene in an organism is already sequenced, then what is the point of genetic mapping? The answer is that a gene's s

Medication Administration - Jones & Bartlett Learning
administration error. Medication error. Oral. Subcutaneous. Sublingual. Transdermal. Transmucosal. CHAPTER OBJECTIVES. A

electrical training - Jones & Bartlett Learning
electronic, printable PDF format. All you need is a computer and the free version of Adobe Reader ..... methods used in

Statistical Significance - Jones & Bartlett Learning
P < 0.05 and distinguish between the statistical significance of a result and its importance in clinical application. 25

Front Matter - Jones & Bartlett Learning
Nov 17, 2015 - facturer, or otherwise does not constitute or imply its endorsement or recommendation by Jones & Bartlett

Chapter 1 - Jones & Bartlett Learning
Describe health communication as used by several government and international organizations. • Understand the logic an

Nutrition Research - Jones & Bartlett Learning
This is the first nutrition research text that starts with the basics and is comprehensive in approach, making it ideal

Pulmonary Ventilation - Jones & Bartlett Learning
By studying this chapter, you should be able to do the following: 1. Identify the basic structures of the ... piration a